PC 50 VioLab bench-top multiparameter with 2301T cell and 201T DHS electrode

Item: 50103892


Professional bench multiparameter with innovative high definition color LCD display, with possibility of brightness adjustment.
Measurement of pH, mV, ORP, Conductivity, TDS, with temperature display.
Possibility to select the parameters you want to see on the display.
The online guide in setup and calibration allows a simple and immediate use of the device.
Automatic pH calibration with USA and NIST buffers up to 3 points and 2 points defined by the user.
Automatic ORP calibration 1 point.
Automatic calibration of conductivity up to 5 points and a user-defined point.
Indication with icons of the calibrated points.
Measurement stability indicator and possibility to select 3 stability levels.
All operations are constantly monitored and signaled to the operator via the colored LED above the display and the self-diagnosis messages.

Measuring range:
pH: 0.00 ... 14.00 (0.01 pH)
mV: -1000 ... + 1000 (0.1 mV)
Cond: 0.0 µS ... 200 mS (0.01 µS - automatic)
TDS: 0.01 mg / l ... 200 g / l
Temperature: 0 ... 100.0 ° C (0.1 ° C)

Supplied as standard:
Electrode holder stand, colored buffer solutions, power supply and instructions for use.

With 201T DHS electrode and 2301T cell, both with built-in temperature sensor and 1 m fixed cable.

Data sheet

Regulations

Regulations

CE

Performance

Accuracy (Cond)

± 2 % f.s.

Accuracy (pH)

± 0,02

Accuracy (Temperature)

± 0,5 °C

Cond measuring range (max.)

200 mS

Cond measuring range (min.)

0,00 µS

Cond resolution

0,01 / 0,1 / 1 µS/cm (automatic scale)

ORP measuring range (max.)

+ 1000

ORP measuring range (min.)

- 1000

ORP resolution

1 mV

pH measuring range (max.)

+ 14,00

pH measuring range (min.)

0

pH resolution

0,01

Reference temperature (max.)

+ 30 °C

Reference temperature (min.)

+ 15 °C

TDS Factor (max)

1.00

TDS Factor (min)

0.40

TDS measuring range (max.)

200 g/l

TDS measuring range (min.)

0,1 mg/l

Temperature coefficent (max)

10,00 % / °C

Temperature coefficent (min)

0,00 % / °C

Temperature compensation

0,00...80,0 °C

Temperature measuring range (max.)

+ 100,0 °C

Temperature measuring range (min.)

0,0 °C

Temperature resolution

0,1 °C

Working temperature (max.)

+ 45 °C

Working temperature (min.)

0 °C

Feature

Calibration report

Yes

DHS compatible

Yes

Display

Backlight coloured high definition LCD

Display

Backlight coloured high definition LCD

Inputs

BNC - RCA

IP protection

IP 54

Points of calibration Cond

1...4

Points of calibration ORP

1

Points of calibration pH

1...3 auto - 2 user values

Printer connection

Yes

Selectable Cell constant

0,1 / 1 / 10

Stability criteria

3 levels

Standard approved Cond

84 µS - 147 µS - 1413 µS - 12,88 mS - 111,8 mS - 1 user value

Standard approved pH

USA - NIST - 2 custom values

Supply

Power supply

100 - 240 V / 50 - 60 Hz

Dimension and Weight

Dimension (LxWxH)

160 x 140 x 45 mm

Weight

380 g