COND 80+ BASIC with VPT 80/1
Item: 50003422
COND 80+ BASIC
Digital microprocessor bench-top conductivity meter with large backlit color graphic display, resolution 0.01 μS. Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 μS, 12.88 and 111 mS) from 1 to 4 points with calculation of the different cell constants and a user-defined standard. Automatic scale change, automatic and manual temperature compensation (0 ... 100.0 ° C) with NT 55 probe. Reference temperature (15 ... 30 ° C), Linear temperature coefficient (0.00 ... 10.00%) and non-linear for ultra pure water. Selectable cell constant (C = 0.1 - 1-10.0). Display on the display of the standards used for calibration. Manual and automatic Data Logger function. Memory / recall up to 1000 data with date / time according to GLP. Self-diagnosis messages, input for cells with 2 electrodes. Intuitive management of parameters via icons. Italian multilingual menu included. Enter company data, operator ID, sample ID and sensor ID via external keyboard (included). Indication of the stability of the measurement. Recalibration alarm, self-diagnosis messages. Alphanumeric password for calibration management and memory cancellation. 2 USB outputs for external keyboard connection (included) and PC. RS232 port for printer connection. Magnetic stirrer with independent control (optional). Quick manual online. PC software downloadable for free. IP54 keyboard, warranty on the electronic part of 3 years.
Measuring range: Conductivity: 0.00 μS ... 1000.0 mS; Resistivity: 1 O ... 10 MW; Salinity: 0.01 ... 100.0 ppt; TDS: 0.1 mg / l ... 500.0 g / l; °C -20.0 to 120.0 (0.1 ° C);
Complete with conductivity cell VPT 1 Platinum glass with 2 electrodes (C = 1) with built-in temperature probe, electrode holder, external keyboard, 2 calibration standards and power supply. Without magnetic stirrer.
Data sheet
Regulations
Regulations
CE
Performance
Accuracy (Cond)
± 1 % f.s.
Accuracy (Resistivity)
± 1 % f.s.
Accuracy (Salinity)
± 1 % f.s.
Accuracy (TDS)
± 1 % f.s.
Accuracy (Temperature)
± 0,2 °C
Cond measuring range (max.)
1000 mS
Cond measuring range (min.)
0,00 µS
Cond resolution
0,01 / 0,1 / 1 µS/cm (automatic scale)
Reference temperature (max.)
+ 30 °C
Reference temperature (min.)
+ 15 °C
Resistivity measuring range (max.)
10,00 MOhm*cm
Resistivity measuring range (min.)
1,00 Ohm*cm
Salinity measuring range (max.)
100 ppt
Salinity measuring range (min.)
0,01 ppm
TDS Factor (max)
1.00
TDS Factor (min)
0.40
TDS measuring range (max.)
500 g/l
TDS measuring range (min.)
0.1 mg/l
Temperature coefficent (max)
10,00 % / °C
Temperature coefficent (min)
0,00 % / °C
Temperature compensation
0,0...100,0 °C
Temperature measuring range (max.)
+ 120,0 °C
Temperature measuring range (min.)
- 20,0 °C
Temperature resolution
0,1 °C
Working temperature (max.)
+ 45 °C
Working temperature (min.)
0 °C
Feature
Points of calibration Cond
1...4
Calibration report
Yes
Cell type
2 poles
Display
Backlight coloured high definition LCD
Display
Backlight coloured high definition LCD
GLP
Yes
GLP
Yes
Inputs
BNC - RCA
IP protection
IP 54
Memory
1000 data
Multilanguage
Eng - Ita - Deu - Esp - Fra - Cze
Online help
Yes
Output
USB - USB type B - RS232
Output
USB - USB type B - RS232
Password
Yes, 4 alphanumerical scripts
Password
Yes, 4 alphanumerical scripts
PC connection
Yes
Printer connection
Yes
Selectable Cell constant
0,1 / 1 / 10
Standard approved Cond
84 µS - 1413 µS - 12,88 mS - 111,8 mS - 1 user value
Timer calibration due
Yes
Ultrapure water compensation
Yes
Supply
Power supply
100 - 240 V / 50 - 60 Hz
Dimension and Weight
Dimension (instrument)
162 x 185 x 56 mm
Instrument weight
709 g