COND 80+ BASIC with VPT 80/1

Item: 50003422


COND 80+ BASIC

Digital microprocessor bench-top conductivity meter with large backlit color graphic display, resolution 0.01 μS. Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 μS, 12.88 and 111 mS) from 1 to 4 points with calculation of the different cell constants and a user-defined standard. Automatic scale change, automatic and manual temperature compensation (0 ... 100.0 ° C) with NT 55 probe. Reference temperature (15 ... 30 ° C), Linear temperature coefficient (0.00 ... 10.00%) and non-linear for ultra pure water. Selectable cell constant (C = 0.1 - 1-10.0). Display on the display of the standards used for calibration. Manual and automatic Data Logger function. Memory / recall up to 1000 data with date / time according to GLP. Self-diagnosis messages, input for cells with 2 electrodes. Intuitive management of parameters via icons. Italian multilingual menu included. Enter company data, operator ID, sample ID and sensor ID via external keyboard (included). Indication of the stability of the measurement. Recalibration alarm, self-diagnosis messages. Alphanumeric password for calibration management and memory cancellation. 2 USB outputs for external keyboard connection (included) and PC. RS232 port for printer connection. Magnetic stirrer with independent control (optional). Quick manual online. PC software downloadable for free. IP54 keyboard, warranty on the electronic part of 3 years.

Measuring range: Conductivity: 0.00 μS ... 1000.0 mS; Resistivity: 1 O ... 10 MW; Salinity: 0.01 ... 100.0 ppt; TDS: 0.1 mg / l ... 500.0 g / l; °C -20.0 to 120.0 (0.1 ° C);

Complete with conductivity cell VPT 1 Platinum glass with 2 electrodes (C = 1) with built-in temperature probe, electrode holder, external keyboard, 2 calibration standards and power supply. Without magnetic stirrer.

Data sheet

Regulations

Regulations

CE

Performance

Accuracy (Cond)

± 1 % f.s.

Accuracy (Resistivity)

± 1 % f.s.

Accuracy (Salinity)

± 1 % f.s.

Accuracy (TDS)

± 1 % f.s.

Accuracy (Temperature)

± 0,2 °C

Cond measuring range (max.)

1000 mS

Cond measuring range (min.)

0,00 µS

Cond resolution

0,01 / 0,1 / 1 µS/cm (automatic scale)

Reference temperature (max.)

+ 30 °C

Reference temperature (min.)

+ 15 °C

Resistivity measuring range (max.)

10,00 MOhm*cm

Resistivity measuring range (min.)

1,00 Ohm*cm

Salinity measuring range (max.)

100 ppt

Salinity measuring range (min.)

0,01 ppm

TDS Factor (max)

1.00

TDS Factor (min)

0.40

TDS measuring range (max.)

500 g/l

TDS measuring range (min.)

0.1 mg/l

Temperature coefficent (max)

10,00 % / °C

Temperature coefficent (min)

0,00 % / °C

Temperature compensation

0,0...100,0 °C

Temperature measuring range (max.)

+ 120,0 °C

Temperature measuring range (min.)

- 20,0 °C

Temperature resolution

0,1 °C

Working temperature (max.)

+ 45 °C

Working temperature (min.)

0 °C

Feature

Points of calibration Cond

1...4

Calibration report

Yes

Cell type

2 poles

Display

Backlight coloured high definition LCD

Display

Backlight coloured high definition LCD

GLP

Yes

GLP

Yes

Inputs

BNC - RCA

IP protection

IP 54

Memory

1000 data

Multilanguage

Eng - Ita - Deu - Esp - Fra - Cze

Online help

Yes

Output

USB - USB type B - RS232

Output

USB - USB type B - RS232

Password

Yes, 4 alphanumerical scripts

Password

Yes, 4 alphanumerical scripts

PC connection

Yes

Printer connection

Yes

Selectable Cell constant

0,1 / 1 / 10

Standard approved Cond

84 µS - 1413 µS - 12,88 mS - 111,8 mS - 1 user value

Timer calibration due

Yes

Ultrapure water compensation

Yes

Supply

Power supply

100 - 240 V / 50 - 60 Hz

Dimension and Weight

Dimension (instrument)

162 x 185 x 56 mm

Instrument weight

709 g