CONDUCTIVITY METERS

The conductivity meters are digital instruments useful to measure electrical conductivity in liquids.
The conductivity is a very important parameter in water analysis.

Conductivity meters are used in Chemical industry, laboratory, typography, school and University, waste water treatment, cooling towers and boilers.

XS®conductivity meters include a full range of bench and portable instruments for measurement in laboratory and field applications.

The new 7-70 series includes a large display ( backlight for 70 serie ), simultaneous indication of conductivity and temperature, indication of stability and calibration data. 
All instruments have an IP 57 certification to guarantee a good protection on field applications.

Conductivity probes and measurement of conductivity :

The measurement of conductivity is done through a probe called conductivity cell that connected to the conductivimeter, plunged in the sample, the value measured is in uS/cm or mS/cm.

There are different cells for different range of measurements:
Cell with constant C=1, range of measurement 10 uS … 10 mS
Cell with constant C=0,1, range of measurement 0,1 uS … 1 mS
Cell with constant C=10, range of measurement 100 uS … 200 mS

Like every physical parameter also the conductivity is influenced by the temperature variation so for this reason many conductivity cell include inside temperature sensor to compensate this variation.

Thanks to temperature compensation it’s possible to give a conductivity value at only one temperature of referement ( normally 20°C or 25°C) independent by real temperature of the sample.

Benchmeters

Cond 51+ With Cell

COND 51+

Conductivity benchtop meter digital microprocessor, with LCD backlight. Simultaneous viewing of Conductivity/TDS, Temperature and values of buffers used for calibration. "Stability" function. Automatic change of scale and offset of temperature automatic and manual (0 ... 80 °C). Selectable Temperature reference (15...30 C) and cell costant (C=0,1; 1; 10). Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 mS and 12,88; 111 mS) and 1 value defined by the user. Messages of self-diagnosis. 3 years warranty on the electronic parts.

Measuring range: Conductivity (Cell with C = 1 ) 0,0 ... 199,9 mS  / TDS: 0...100 g/l  /  °C 0,0...100,0 (0,1 °C ).

Complete of cell 2301T (C = 1) 2-electrode with integrated temperature sensor, support electrode, standard solutions and user manual.

Cond 51+ Without Cell

COND 51+

Conductivity benchtop meter digital microprocessor, with LCD backlight. Simultaneous viewing of Conductivity/TDS, Temperature and values of buffers used for calibration. "Stability" function. Automatic change of scale and offset of temperature automatic and manual (0 ... 80 °C). Selectable Temperature reference (15...30 °C) and cell costant (C=0,1; 1; 10). Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 mS and 12,88; 111 mS) and 1 value defined by the user. Messages of self-diagnosis. 3 years warranty on the electronic parts.

Measuring rangeConductivity (Cell with C = 1 ) 0,0 ... 199,9 mS  / TDS0...100 g/l  / °C 0,0...100,0 (0,1 °C ).

Without cell 2301T (C = 1) but with support electrode, standard solutions and instructions for use.

Cond 61 with Cell

COND 61

Conductivity benchtop meter digital microprocessor, LCD backlight display with 3 colours with simultaneous viewing of Conductivity/TDS, Temperature and values of buffers used for calibration. GLP function with last calibration's date and hour. "CAL TIMER" function for calibration setting frequency. "Stability" function. Automatic change of scale and offset of temperature automatic and manual (0 ... 100 ° C). Selectable Temperature reference (15...30 C) and cell costant (C=0,1; 1; 10). Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 mS and 12,88; 111 mS) and 1 value defined by the user. Data Logger automatic or manual up to 500 measurements with date and hour. USB door for PC and RS232 for printer. Messages of self-diagnosis. 3 years warranty on the electronic parts.

Measuring rangeConductivity (Cell with C = 1 ) 0,0 ... 199,9 mS / TDS0...100 g/l  / Salinity: 0...50 g/l /  °C 0,0...100,0 (0,1 °C ).

With cell 2301T (C=1) 2-electrode with temperature probe, 1 m fixed cable, BNC and Cinch connectors. 220V power supply, PC-Link software with USB cable, buffer solutions and user manual. 

Cond 61 without Cell

COND 61

Conductivity benchtop meter digital microprocessor, LCD backlight display with 3 colours with simultaneous viewing of Conductivity/TDS, Temperature and values of buffers used for calibration. GLP function with last calibration's date and hour. "CAL TIMER" function for calibration setting frequency. "Stability" function. Automatic change of scale and offset of temperature automatic and manual (0 ... 100 °C). Selectable Temperature reference (15...30 °C) and cell costant (C=0,1; 1; 10). Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 mS and 12,88; 111 mS) and 1 value defined by the user. Data Logger automatic or manual up to 500 measurements with date and hour. USB door for PC and RS232 for printer. Messages of self-diagnosis. 3 years warranty on the electronic parts.

Measuring rangeConductivity (Cell with C = 1 ) 0,0 ... 199,9 mS / TDS0...100 g/l  / Salinity: 0...50 g/l /  °C 0,0...100,0 (0,1 °C )..

Without cell but with 220V power supply, RS 232C cable, software, buffer solutions and user manual.

COND 8+ BASIC with 2301T cell

COND 8+ BASIC

Digital microprocessor bench-top conductivity meter with large backlit color graphic display, resolution 0.01 μS. Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 μS, 12.88 and 111 mS) from 1 to 4 points with calculation of the different cell constants and a user-defined standard. Automatic scale change, automatic and manual temperature compensation (0 ... 100.0 ° C) with NT 55 probe. Reference temperature (15 ... 30 ° C), Linear temperature coefficient (0.00 ... 10.00%), Selectable cell constant (C = 0.1 - 1-10.0). Display on the display of the standards used for calibration. Manual and automatic Data Logger function. Memory / recall up to 1000 data with date / time according to GLP. Self-diagnosis messages, input for cells with 2 electrodes. Intuitive management of parameters via icons. Italian multilingual menu included. Indication of the stability of the measurement. Recalibration alarm, self-diagnosis messages. Numerical password for calibration management and memory cancellation. 1 USB output for PC connection. RS232 port for printer connection. Magnetic stirrer with independent control (optional). PC software downloadable for free. IP54 keyboard, warranty on the electronic part of 3 years.

Measurement range: Conductivity: 0.00 μS ... 200.0 mS; Salinity: 0.01 ... 100.0 ppt; TDS: 0.1 mg / l ... 100.0 g / l; °C -10.0 ... 110.0 (0.1 ° C);

Complete with 2301T epoxy conductivity cell with 2 electrodes (C = 1) with built-in temperature probe, electrode holder, 2 calibration standards and power supply. Without agitator.


COND 8 BASIC Without Cell, Without STIRRER

COND 8+ BASIC

Digital microprocessor bench-top conductivity meter with large backlit color graphic displayresolution 0.01 μS. Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 μS, 12.88 and 111 mS) from 1 to 4 points with calculation of the different cell constants and a user-defined standard. Automatic scale change, automatic and manual temperature compensation (0 ... 100.0 ° C) with NT 55 probe. Reference temperature (15 ... 30 ° C), Linear temperature coefficient (0.00 ... 10.00%), Selectable cell constant (C = 0.1 - 1-10.0). Display on the display of the standards used for calibration. Manual and automatic Data Logger function. Memory / recall up to 1000 data with date / time according to GLP. Self-diagnosis messages, input for cells with 2 electrodes. Intuitive management of parameters via icons. Italian multilingual menu included. Indication of the stability of the measurement. Recalibration alarm, self-diagnosis messages. Numerical password for calibration management and memory cancellation. 1 USB output for PC connection. RS232 port for printer connection. Magnetic stirrer with independent control (optional). PC software downloadable for free. IP54 keyboard, warranty on the electronic part of 3 years.

Measurement rangeConductivity: 0.00 μS ... 200.0 mS; Salinity: 0.01 ... 100.0 ppt; TDS: 0.1 mg / l ... 100.0 g / l; °C -10.0 ... 110.0 (0.1 ° C);

Without cell, with electrode holder, 2 calibration standards and power supply. Without agitator.


COND 8+ STIRRER With Cell 2301 T

COND 8+ STIRRER

Digital microprocessor bench-top conductivity meter with large backlit color graphic displayresolution 0.01 μS. Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 μS, 12.88 and 111 mS) from 1 to 4 points with calculation of the different cell constants and a user-defined standard. Automatic scale change, automatic and manual temperature compensation (0 ... 100.0 ° C) with NT 55 probe. Reference temperature (15 ... 30 ° C), Linear temperature coefficient (0.00 ... 10.00%), Selectable cell constant (C = 0.1 - 1-10.0). Display on the display of the standards used for calibration. Manual and automatic Data Logger function. Memory / recall up to 1000 data with date / time according to GLP. Self-diagnosis messages, input for cells with 2 electrodes. Intuitive management of parameters via icons. Italian multilingual menu included. Indication of the stability of the measurement. Recalibration alarm, self-diagnosis messages. Numerical password for calibration management and memory cancellation. 1 USB output for PC connection. RS232 port for printer connection. Magnetic stirrer with independent control (optional). PC software downloadable for free. IP54 keyboard, warranty on the electronic part of 3 years.

Measurement rangeConductivity: 0.00 μS ... 200.0 mS; Salinity: 0.01 ... 100.0 ppt; TDS: 0.1 mg / l ... 100.0 g / l; °C -10.0 ... 110.0 (0.1 ° C);

Complete with 2301T epoxy conductivity cell with 2 electrodes (C = 1) with built-in temperature probe, electrode holder, magnetic stirrer with adjustable speed, 2 calibration standards and power supply. 


COND 8+ STIRRER Without Cell

COND 8+ STIRRER

Digital microprocessor bench-top conductivity meter with large backlit color graphic displayresolution 0.01 μS. Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 μS, 12.88 and 111 mS) from 1 to 4 points with calculation of the different cell constants and a user-defined standard. Automatic scale change, automatic and manual temperature compensation (0 ... 100.0 ° C) with NT 55 probe. Reference temperature (15 ... 30 ° C), Linear temperature coefficient (0.00 ... 10.00%), Selectable cell constant (C = 0.1 - 1-10.0). Display on the display of the standards used for calibration. Manual and automatic Data Logger function. Memory / recall up to 1000 data with date / time according to GLP. Self-diagnosis messages, input for cells with 2 electrodes. Intuitive management of parameters via icons. Italian multilingual menu included. Indication of the stability of the measurement. Recalibration alarm, self-diagnosis messages. Numerical password for calibration management and memory cancellation. 1 USB output for PC connection. RS232 port for printer connection. Magnetic stirrer with independent control (optional). PC software downloadable for free. IP54 keyboard, warranty on the electronic part of 3 years.

Measurement rangeConductivity: 0.00 μS ... 200.0 mS; Salinity: 0.01 ... 100.0 ppt; TDS: 0.1 mg / l ... 100.0 g / l; °C -10.0 ... 110.0 (0.1 ° C);

Without cell, electrode holder, magnetic stirrer with adjustable speed, 2 calibration standards and power supply. 


COND 80+ STIRRER With Cell VPT 80-1

COND 80+ STIRRER

Digital microprocessor bench-top conductivity meter with large backlit color graphic display, resolution 0.01 μS. Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 μS, 12.88 and 111 mS) from 1 to 4 points with calculation of the different cell constants and a user-defined standard. Automatic scale change, automatic and manual temperature compensation (0 ... 100.0 ° C) with NT 55 probe. Reference temperature (15 ... 30 ° C), Linear temperature coefficient (0.00 ... 10.00%) and non-linear for ultra pure water. Selectable cell constant (C = 0.1 - 1-10.0). Display on the display of the standards used for calibration. Manual and automatic Data Logger function. Memory / recall up to 1000 data with date / time according to GLP. Self-diagnosis messages, input for cells with 2 electrodes. Intuitive management of parameters via icons. Italian multilingual menu included. Enter company data, operator ID, sample ID and sensor ID via external keyboard (included). Indication of the stability of the measurement. Recalibration alarm, self-diagnosis messages. Alphanumeric password for calibration management and memory cancellation. 2 USB outputs for external keyboard connection (included) and PC. RS232 port for printer connection. Magnetic stirrer with independent control (optional). Quick manual online. PC software downloadable for free. IP54 keyboard, warranty on the electronic part of 3 years.

Measuring range: Conductivity: 0.00 μS ... 1000.0 mS; Resistivity: 1 O ... 10 MW; Salinity: 0.01 ... 100.0 ppt; TDS: 0.1 mg / l ... 500.0 g / l; °C -20.0 to 120.0 (0.1 ° C);

Complete with conductivity cell VPT 80/1 Platinum glass with 2 electrodes (C = 1) with built-in temperature probe, electrode holder, external keyboard, magnetic stirrer with adjustable speed, 2 calibration standards and power supply. 

COND 80+ STIRRER Without Cell

COND 80+ STIRRER 

Digital microprocessor bench-top conductivity meter with large backlit color graphic display, resolution 0.01 μS. Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 μS, 12.88 and 111 mS) from 1 to 4 points with calculation of the different cell constants and a user-defined standard. Automatic scale change, automatic and manual temperature compensation (0 ... 100.0 ° C) with NT 55 probe. Reference temperature (15 ... 30 ° C), Linear temperature coefficient (0.00 ... 10.00%) and non-linear for ultra pure water. Selectable cell constant (C = 0.1 - 1-10.0). Display on the display of the standards used for calibration. Manual and automatic Data Logger function. Memory / recall up to 1000 data with date / time according to GLP. Self-diagnosis messages, input for cells with 2 electrodes. Intuitive management of parameters via icons. Italian multilingual menu included. Enter company data, operator ID, sample ID and sensor ID via external keyboard (included). Indication of the stability of the measurement. Recalibration alarm, self-diagnosis messages. Alphanumeric password for calibration management and memory cancellation. 2 USB outputs for external keyboard connection (included) and PC. RS232 port for printer connection. Magnetic stirrer with independent control (optional). Quick manual online. PC software downloadable for free. IP54 keyboard, warranty on the electronic part of 3 years.


Measuring range: Conductivity: 0.00 μS ... 1000.0 mS; Resistivity: 1 O ... 10 MW; Salinity: 0.01 ... 100.0 ppt; TDS: 0.1 mg / l ... 500.0 g / l; °C -20.0 to 120.0 (0.1 ° C);

Without conductivity cell, complete with electrode holder support, external keyboard, adjustable speed magnetic stirrer, 2 calibration standards and power supply.

COND 80+ BASIC with VPT 80/1

COND 80+ BASIC

Digital microprocessor bench-top conductivity meter with large backlit color graphic display, resolution 0.01 μS. Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 μS, 12.88 and 111 mS) from 1 to 4 points with calculation of the different cell constants and a user-defined standard. Automatic scale change, automatic and manual temperature compensation (0 ... 100.0 ° C) with NT 55 probe. Reference temperature (15 ... 30 ° C), Linear temperature coefficient (0.00 ... 10.00%) and non-linear for ultra pure water. Selectable cell constant (C = 0.1 - 1-10.0). Display on the display of the standards used for calibration. Manual and automatic Data Logger function. Memory / recall up to 1000 data with date / time according to GLP. Self-diagnosis messages, input for cells with 2 electrodes. Intuitive management of parameters via icons. Italian multilingual menu included. Enter company data, operator ID, sample ID and sensor ID via external keyboard (included). Indication of the stability of the measurement. Recalibration alarm, self-diagnosis messages. Alphanumeric password for calibration management and memory cancellation. 2 USB outputs for external keyboard connection (included) and PC. RS232 port for printer connection. Magnetic stirrer with independent control (optional). Quick manual online. PC software downloadable for free. IP54 keyboard, warranty on the electronic part of 3 years.

Measuring range: Conductivity: 0.00 μS ... 1000.0 mS; Resistivity: 1 O ... 10 MW; Salinity: 0.01 ... 100.0 ppt; TDS: 0.1 mg / l ... 500.0 g / l; °C -20.0 to 120.0 (0.1 ° C);

Complete with conductivity cell VPT 80/1 Platinum glass with 2 electrodes (C = 1) with built-in temperature probe, electrode holder, external keyboard, 2 calibration standards and power supply. Without magnetic stirrer.

COND 80+ BASIC without conductivity cell

COND 80+ BASIC

Digital microprocessor bench-top conductivity meter with large backlit color graphic display, resolution 0.01 μS. Automatic and manual calibration with recognition of 4 conductivity standards (84; 1413 μS, 12.88 and 111 mS) from 1 to 4 points with calculation of the different cell constants and a user-defined standard. Automatic scale change, automatic and manual temperature compensation (0 ... 100.0 ° C) with NT 55 probe. Reference temperature (15 ... 30 ° C), Linear temperature coefficient (0.00 ... 10.00%) and non-linear for ultra pure water. Selectable cell constant (C = 0.1 - 1-10.0). Display on the display of the standards used for calibration. Manual and automatic Data Logger function. Memory / recall up to 1000 data with date / time according to GLP. Self-diagnosis messages, input for cells with 2 electrodes. Intuitive management of parameters via icons. Italian multilingual menu included. Enter company data, operator ID, sample ID and sensor ID via external keyboard (included). Indication of the stability of the measurement. Recalibration alarm, self-diagnosis messages. Alphanumeric password for calibration management and memory cancellation. 2 USB outputs for external keyboard connection (included) and PC. RS232 port for printer connection. Magnetic stirrer with independent control (optional). Quick manual online. PC software downloadable for free. IP54 keyboard, warranty on the electronic part of 3 years.

Measuring range: Conductivity: 0.00 μS ... 1000.0 mS; Resistivity: 1 O ... 10 MW; Salinity: 0.01 ... 100.0 ppt; TDS: 0.1 mg / l ... 500.0 g / l; °C -20.0 to 120.0 (0.1 ° C);

Without conductivity cell, complete with electrode holder stand, external keyboard, 2 calibration standards and power supply. Without magnetic stirrer.

VPT80-1 Cell Glass-Pt C=1+ATC

VPT 80/1 Platinum glass conductivity cell C = 1 with built-in temperature sensor for all series 7 conductivity meters; 70; 51; 8; 80. Field 10μS ... 100 mS.

Cell 2301 T C=1 with ATC for COND 51

2301T standard spare cell C = 1 for COND 51 conductivity meter, with 2 electrodes with built-in temperature sensor, epoxy body, 1 meter cable. Working range 10μS ... 200 mS.

Cell VPT51-01 C=0,1 for COND 7*70*51

VPT 51/01 steel / PVC cell with built-in temperature sensor, C = 0.1, 1m fixed cable with BNC and Cinch connector Range 0.1μS ... 1 mS.

VPT80-10 Cell Glass-PT C=10+ATC

VPT 80/10 Platinum glass conductivity cell C=10 with built-in temperature sensor, 1m fixed cable with BNC and Cinch connector. Field 100μS ... 200mS.

Portable

COND 7+ complete instrument WITH CELL 2301T+ ATC

COND 7+

Conductivity portable meter with microprocessor. Simultaneous display of conductivity/TDS, Temperature and the values of buffers used for calibration. Function "Stability". Automatic scale change and automatic temperature compensation and manual (0 ... 80 °C). Adjustable reference temperature (15 ... 30 °C) and the cell constant (C = 0.1; 1; 10) selectable. Automatic and manual calibration with automatic recognition of 4 standard (84; 1413 µS and 12,88; 111 mS) and one user-defined value. Auto-off, messages of self-diagnosis. Rubber cover with bench support. 5-year warranty on the electronics. IP57 Waterproof.

Measuring range: (cell with C = 1 ) 0.0 ... 199.9 mS / TDS: 0...100 g/l  /  °C 0.0 ... 100.0 (0.1 °C)

Complete cell 2301T 2-electrode with integrated temperature sensor, standard solutions and instructions for use, in carrying case.

COND 7+TDS  INSTRUMENT without CELL

COND 7+ 

Conductivity portable meter with microprocessor. Simultaneous display of conductivity/TDS, Temperature and the values of buffers used for calibration. Function "Stability". Automatic scale change and automatic temperature compensation and manual (0 ... 80 °C). Adjustable reference temperature (15 ... 30 °C) and the cell constant (C = 0.1; 1; 10) selectable. Automatic and manual calibration with automatic recognition of 4 standard (84; 1413 µS and 12,88; 111 mS) and one user-defined value. Auto-off, messages of self-diagnosis. Rubber cover with bench support. 5-year warranty on the electronics. IP57 Waterproof.

Measuring range: (cell with C = 1 ) 0.0 ... 199.9 mS / TDS: 0...100 g/l  /  °C 0.0 ... 100.0 (0.1 °C)

Without cell 2301T but with standard solutions and instructions for use, in carrying case.

COND 70+ complete instrument WITH CELL 2301T+ ATC

COND 70+

Conductivity portable meter with microprocessor and triple power (via PC, network and battery) output with USB for PC connection. Large backlight display with 3 colours with simultaneous display of conductivity/TDS, temperature and values of buffers used for calibration. GLP with date and time of last calibration. Function "CAL TIMER" for setting the frequency of recalibration. Function "Stability". Automatic scale change and automatic temperature compensation and manual (0.0 ... 100 °C). The reference temperature (15 ... 30 ° C) and the cell constant (C = 0.1; 1; 10) selectable. Automatic and manual calibration with automatic recognition of 4 standard (84; 1413 µS and 12,88; 111 mS) and one user-defined value. Data Logger with manual or automatic memory / recall up to 500 data with date and time. Rubber cover with bench support. Messages of self-diagnosis, Auto-Off . 5 year warranty on the electronics . IP57 Waterproof.

Measuring range: (cell with C = 1 ) Conductivity: 0 ... 199.9 mS / TDS: 0 ... 100 g/l / Salinity: 0 ... 50g/l / °C: -10.0 ... 110.0 °C

Complete cell 2301T 2-electrode with integrated temperature sensor, 1 m fixed cable, BNC plug and Cinch connector for temperature sensor, AC adapter 220V, PC-Link software with USB cable for data download to PC, buffers colourful and instructions for use, in carrying case.

COND 70+  INSTRUMENT without CELL

COND 70+

Conductivity portable meter with microprocessor and triple power (via PC, network and battery) output with USB for PC connection. Large backlight display with 3 colours with simultaneous display of conductivity/TDS, temperature and values of buffers used for calibration. GLP with date and time of last calibration. Function "CAL TIMER" for setting the frequency of recalibration. Function "Stability". Automatic scale change and automatic temperature compensation and manual (0.0 ... 100 ° C). The reference temperature (15 ... 30 ° C) and the cell constant (C = 0.1; 1; 10) selectable. Automatic and manual calibration with automatic recognition of 4 standard (84; 1413 µS and 12,88; 111 mS) and one user-defined value. Data Logger with manual or automatic memory / recall up to 500 data with date and time. Rubber cover with bench support. Messages of self-diagnosis, Auto-Off . 5 year warranty on the electronics . IP57 Waterproof.

Measuring range: (cell with C = 1 ) Conductivity: 0 ... 199.9 mS / TDS: 0 ... 100 g/l / Salinity: 0 ... 50g/l / °C: -10.0 ... 110.0 °C

Without cell 2301T, but with 220V AC adapter, PC-Link software with USB cable for data download to PC, buffers colourful and instructions for use, in carrying case.

Cell 2301 T C=1 with ATC 5MT

2301T standard spare cell C = 1 for COND 51 conductivity meter, with 2 electrodes with built-in temperature sensor, epoxy body, 5 meter cable. Working range 10μS ... 200 mS.

VPT SOIL conductivity cell for series 7 and 70

VPT Soil conductivity cell with built-in temperature sensor, C = 1, for ground measurement, 1.5mt fixed cable with BNC and Cinch connector. Field 10μS ... 50 mS. For Cond 7 - 70